The customer reported a microclave neutral connector that leaked roughly 50-75mls of blood on a patient with low platelets.The event was discovered during a routine hourly access check.Intravenous fluids were reported as being used.The medical team ordered antibiotics.The set up was a huber needle to clave.There was patient involvement, but no adverse event, or human harm reported.
|
H10: one (1) used list# b3300, clave¿ neutral connector.Lot# unknown, one (1) used spinning spiros, unknown list#, unknown lot#, one (1) extension set, unknown manufacturer along with one (1) safestep huber needle set, lot number ascv0202, ref 0724134.Manufacturer bard were received on august 31, 2020 for evaluation.The b3300 clave neutral connector was pressure leak tested as returned with the bard safestep huber needle set as described in the complaint and there was no leakage or propensity to disconnect observed during testing.The luers of the b3300 clave neutral connector and bard safestep huber needle set were both measured and both were found to be iso compatible.The complaint of leakage was unable to be replicated or confirmed.A device history review lot review could not be conducted because no lot number(s) was/were identified.
|