• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Device Problem Structural Problem (2506)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Follow-up #1 date of submission 03/23/2016 device evaluation: the pump has been returned and evaluated by product analysis on 03/14/2016 with the following findings: a visual inspection of the pump showed that the battery compartment was cracked.Unrelated to the complaint, the audio bolus button cover was torn.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Event Description
On (b)(6) 2016, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.It was reported that the battery compartment was damaged.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Type of Device
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer (Section G)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
Manufacturer Contact
jon mulberg
965 chesterbrook blvd
wayne, PA 19087
MDR Report Key10528115
MDR Text Key207377451
Report Number2531779-2020-00256
Device Sequence Number1
Product Code OYC
Combination Product (y/n)Y
PMA/PMN Number
P130007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2016
Was the Report Sent to FDA? Yes
Device Age33 MO
Date Manufacturer Received03/14/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-