Related manufacturer reference number: 2017865-2020-12508.It was reported that the left ventricular (lv) lead exhibited high pacing impedance and loss of capture.During lead revision procedure, it was noted that the set screw on the implantable cardioverter defibrillator (icd) would not tighten after the new lead was placed, it was suspected that the lv header on the device could have resulted in the anomalies.The icd and lv lead were explanted and replaced.The patient was in stable condition before, during, and after the procedure.
|