Model Number 2090
Device Problems Computer Software Problem (1112); Data Problem (3196)
Patient Problem No Patient Involvement (2645)
Event Date 07/28/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was unable to reproduce the black screen error however a system error was recorded twice in the recent log file - the hard drive was reconfigured.Inspected and tested all three usb ports and no anomalies were observed.Inspected nylon standoff between link electronic module (lem) and micro processor unit (mpu) and no anomalies were noticed - replaced as a preventative.Power cord was missing - power cord was added.Inspected circuit boards and mechanical parts.Programmer failed thermal sensor 7 - the power supply was replaced.The software was reloaded and updated and the device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer displayed a black screen error and there was a request to check the universal serial bus (usb) port function.The programmer was returned for service.It was further reported that the programmer subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10533952
MDR Text Key206933199
Report Number2182208-2020-01865
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994581570
UDI-Public00613994581570
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2020
Date Manufacturer Received08/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2067 RADIOFREQUENCY HEAD.
-
-