Model Number 21-7302-24 |
Device Problems
Device Alarm System (1012); Pressure Problem (3012)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/05/2020 |
Event Type
malfunction
|
Event Description
|
Information was received indicating that during use of a smiths medical cadd cassette reservoir, high pressure message was displayed on the cadd infusion pump.It was also reported that the alarm endured until cassette was changed.No patient consequences were reported.No adverse effects were reported.
|
|
Manufacturer Narrative
|
Other, other text: 11 units of cadd cassette reservoirs were returned to investigate the reported event of causing pumps to alarm.The units were visually inspected at a distance of 12" to 16" under normal conditions of illumination.No obstructions nor occlusion were detected in any of the cassette received.A functional test was also performed to look for unusual functions, but the reported issue could not be replicated.Root cause could not be determined since the complaint was not confirmed due to the fact the sample were tested and the alarm was not activated.No further actions were taken.
|
|
Manufacturer Narrative
|
Returned device was received for evaluation.During the evaluation of the device the customer reported condition was not confirmed.No fault found.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.
|
|
Search Alerts/Recalls
|