The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of a damaged obturator tip.The tip of the obturator was fractured; however, the fragment was not returned for evaluation.Based on the condition of the returned unit, it is likely that the obturator tip contained a material abnormality, which could have made it more susceptible to damage.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.A bent obturator tip is not considered to be reportable as it is unlikely to cause or contribute to death or serious injury.However, the event is reportable due to the fractured obturator tip that was observed during the evaluation of the returned unit.
|