MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 SYSTEM INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9735669
Device Problems Connection Problem (2900); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/11/2020
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable at the time of filing. Other relevant device(s) are: product id: 9733752, serial/lot #: unknown. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used for a functional endoscopic sinus surgery (fess) procedure. It was reported that the system showed "localizer faulted". The site attempted a reboot and it did not resolve the issue. Further troubleshooting was performed by disconnecting and reconnecting the emitter which did not resolve the issue. The site also disconnected the breakout box and this did not resolve the issue. The on-screen messages changed while each component was unplugged, but returned to "localizer faulted" when the emitter was plugged in. The site was using a flat panel emitter. Navigation was aborted to complete the procedure. There was a less than 1 hour delay and no impact to patient outcome as a result of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S8 SYSTEM
Type of DeviceINSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key10537602
MDR Text Key207162058
Report Number1723170-2020-02456
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169838918
UDI-Public00643169838918
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 11/10/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735669
Device Catalogue Number9735669
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/22/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-