• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TRACKING CAMERA; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE TRACKING CAMERA; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB10025
Device Problem Intermittent Communication Failure (4038)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/17/2020
Event Type  malfunction  
Event Description
It was reported that during a cori ukr procedure, the camera was disconnected because it was blocked by the surgeon and it would not reconnect quickly.They restarted the system and system indicate that both femur and tibia tracker were moved ("confirm redefinition").However, it is rarely that both femur and tibia tracker moved.Although this happened after trial reduction and did not extend the surgical time.They shut down the system after the case and there was a warming box showing system fault and critical error indicating that the launcher exited due to configuration error.No other complications were reported.
 
Manufacturer Narrative
The real intelligence tracking camera, part number rob10025, serial unknown and used for treatment, was not returned for evaluation.A relationship between the reported event and the device could not be established.A visual/functional inspection cannot be completed as no device was returned for evaluation.While all products meet required manufacturing specifications prior to release a serial number or lot number is required to link the device to a dhr or nc investigation, or field service report.A complaint history review for similar reported/confirmed complaints has identified prior events.A contributing factor could be software issues.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE TRACKING CAMERA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10538629
MDR Text Key207318744
Report Number3010266064-2020-01762
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB10025
Was Device Available for Evaluation? No
Date Manufacturer Received09/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-