• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA 100; LAMP, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA 100; LAMP, SURGICAL Back to Search Results
Catalog Number ARD569070999
Device Problem Peeled/Delaminated (1454)
Patient Problem No Patient Involvement (2645)
Event Date 08/25/2020
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided upon results of investigation.Device not returned to manufacturer.
 
Event Description
On 25th of august 2020 getinge became aware if an issue related to one of our device - lucea 100.As stated customer alleged cracks on near to the switch of lucea device and during the device inspection paint chipping was discovered.No information abour patient involvement was reported however we decided to report this case in abundance of caution and based on potential as any paint particles falling into sterile field might be a source of contamination.Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event Description
Manufacturer reference nmber (b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
Getinge became aware if an issue related to one of our device - lucea 100.As stated customer alleged cracks on near to the switch of lucea device and during the device inspection paint chipping was discovered.No information about patient involvement was reported however we decided to report this case in abundance of caution and based on potential as any paint particles falling into sterile field might be a source of contamination.Technician provided information that new parts were ordered.It was established that when the event occurred, the surgical light did not meet its specification as particles of paint were missing and cover was cracked, and it contributed to event.None of the provided information indicate that upon the event occurrence the device was being used for patient treatment.When reviewing similar reportable events for the same device type, we have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.All maquet sas products comply with: ¿ iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.¿ iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.¿ paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.¿ disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.Cover cracking is indicated by our product experts to likely be caused by combination of different factors such as: mechanical stress, environmental conditions (such a high temperature and humidity variations during transport and storage), use of inappropriate cleaning/disinfections products, or inappropriate cleaning and disinfection protocols.To prevent any similar incident, it is recommended to avoid the collisions between devices.Visual inspections during the cleaning allow detecting the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor pain chip can be repaired with touch up paint; nevertheless, the parts impacted by serious damage must be replaced.We believe that overall the devices on the market are performing correctly.Given the circumstances we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 100
Type of Device
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key10539420
MDR Text Key207122776
Report Number9710055-2020-00367
Device Sequence Number1
Product Code FTD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberARD569070999
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-