Getinge became aware if an issue related to one of our device - lucea 100.As stated customer alleged cracks on near to the switch of lucea device and during the device inspection paint chipping was discovered.No information about patient involvement was reported however we decided to report this case in abundance of caution and based on potential as any paint particles falling into sterile field might be a source of contamination.Technician provided information that new parts were ordered.It was established that when the event occurred, the surgical light did not meet its specification as particles of paint were missing and cover was cracked, and it contributed to event.None of the provided information indicate that upon the event occurrence the device was being used for patient treatment.When reviewing similar reportable events for the same device type, we have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.All maquet sas products comply with: ¿ iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.¿ iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.¿ paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.¿ disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.Cover cracking is indicated by our product experts to likely be caused by combination of different factors such as: mechanical stress, environmental conditions (such a high temperature and humidity variations during transport and storage), use of inappropriate cleaning/disinfections products, or inappropriate cleaning and disinfection protocols.To prevent any similar incident, it is recommended to avoid the collisions between devices.Visual inspections during the cleaning allow detecting the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor pain chip can be repaired with touch up paint; nevertheless, the parts impacted by serious damage must be replaced.We believe that overall the devices on the market are performing correctly.Given the circumstances we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
|