Model Number 6935M55
Device Problems Over-Sensing (1438); Under-Sensing (1661); High Capture Threshold (3266)
Patient Problem Palpitations (2467)
Event Date 08/27/2020
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: 5076-45, lead, implanted: (b)(6) 2019.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the right ventricular (rv) lead exhibited high pacing thresholds which coincided with short ventricular intervals.The patient reportedly experienced a pounding sensation in their chest.The lead remains in use.No further patient complications have been reported as a result of this event.
 
Event Description
It was further reported that a lead integrity alert (lia) was triggered and t-wave oversensing (twos) was observed on the right ventricular (rv) lead.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated pacing capture threshold in the right ventricle was elevated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that rv lead exhibited occasional dropped beats.The rv lead was reprogrammed and remains in use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key10539970
MDR Text Key207116320
Report Number2649622-2020-17738
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00643169356559
UDI-Public00643169356559
Combination Product (y/n)N
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/16/2022
Device Model Number6935M55
Device Catalogue Number6935M55
Was Device Available for Evaluation? No
Date Manufacturer Received10/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
DTMB1QQ CRT-D, 459888 LEAD; DTMB1QQ CRT-D, 459888 LEAD
Patient Outcome(s) Required Intervention;
Patient Age59 YR
Patient Weight73
-
-