Model Number 3L92510 |
Device Problems
Fracture (1260); Off-Label Use (1494)
|
Patient Problem
No Code Available (3191)
|
Event Date 08/28/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Revision of a femoral stem that broken inside the patient.The primary surgery was in 2012.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.The concomitant product was identified to be a competitor product used in conjunction with the previously reported adverse event.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Additional information received stated that the affected side is the left side.Surgery time was not extended.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Review of the attached x-ray image confirmed the reported allegation.The root cause could not be determined.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : lot 3453095.Dhr has been reviewed and no deviations or anomalies were found.
|
|
Search Alerts/Recalls
|