• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Loose or Intermittent Connection (1371); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 08/25/2020
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue. The pressure port was not making a solid connection after being cleaned. The iabp port was not able to make connection using mini sim connection. To fix the issue, the fse replaced the assembly external connection as required, and performed all functional and safety checks to meet factory specifications. Unit passed all functional and safety test per factory specifications. The iabp was then released to the customer and cleared for clinical service. Full event site name: (b)(6).
 
Event Description
It was reported that during a check performed by the customer, the pressure port of the cs300 intra-aortic balloon pump (iabp) was not making a good connection with the customer's slave cables, the port was loose. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS300
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key10541531
MDR Text Key207192551
Report Number2249723-2020-01488
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 10/06/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-