Model Number 100014514
Device Problem Failure of Device to Self-Test (2937)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/16/2020
Event Type  Injury  
Manufacturer Narrative
The results, method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During an ablation procedure, while ablation was being performed the amplifier status light turned orange and an amplifier error message was received. The amplifier was power cycled and the tactisys and generator to amplifier cables were changed but the issue remained. As ablation was unable to be continued, the procedure was cancelled. There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameENSITE VELOCITY SYSTEM VELOCITY AMPLIFIER
Type of DeviceCOMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key10541571
MDR Text Key207172917
Report Number2184149-2020-00148
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K141050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/28/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number100014514
Device Catalogue Number100014514
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-