• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL HD EPSCP,4.0,30,167,MITEK; RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL HD EPSCP,4.0,30,167,MITEK; RIGID ENDOSCOPE Back to Search Results
Model Number 242018
Device Problems Poor Quality Image (1408); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/27/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: the device was received and evaluated at the service center.The reported complaint that the scope was cracked, was confirmed.The following defects were found with the device upon evaluation by the manufacturer: outer tube damaged, distal tip damaged, shaver damage to distal tip, distal tip has deposits, image error, on camera, image cloudy/blurred.The device was diagnosed and repaired using spare parts.It was tested and found to be working according to specifications.User mishandling of the device, probably due to a fall, is the most probable root cause of the physical damages to the device.The distal tip has been damaged due to contact with the shaver during a surgical procedure as identified during evaluation.Also the deposits on the distal tip can be attributed to lack of cleaning and maintenance of the device by the customer.There are no indications from this complaint investigation that the failures are manufacturing-related, therefore a manufacturing record evaluation is not required.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Event Description
It was reported by affiliate via email that during procedure the hd epscp, 4.0, 30, 167, mitek was cracked.No surgical delay or patient consequence reported.The device is available to be returned for evaluation.Additional information provided by the affiliate reported the failure was noted during pre-operative activities.It was reported the case was completed with another readily available device and surgical intervention is not required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD EPSCP,4.0,30,167,MITEK
Type of Device
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key10542297
MDR Text Key207219035
Report Number1221934-2020-02637
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705029150
UDI-Public10886705029150
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080560
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 09/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number242018
Device Catalogue Number242018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2020
Date Manufacturer Received09/11/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-