Product complaint #: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the device did not turn and the motor was jammed.Per service reports, this complaint can be confirmed.It was found during evaluation that the motor was rusty and stuck.Further, the cable resistance was out of tolerance and the o-rings were defective.The motor, motor cable and o-rings were replaced to resolve the issues.The device was cleaned, tested, and found to be fully functional.Fluid ingress into the device and contact with the motor is responsible for the rusty motor.Corroded motor has a tendency to stick and not turn.With the available information, we cannot determine the root cause of the other identified failures.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 02/13/2019 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|