Model Number AB-5100L |
Device Problems
Electrical /Electronic Property Problem (1198); Adverse Event Without Identified Device or Use Problem (2993); Intermittent Communication Failure (4038)
|
Patient Problem
Hearing Impairment (1881)
|
Event Date 08/21/2020 |
Event Type
Injury
|
Event Description
|
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issue did not resolve.Revision surgery is under consideration.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.The recipient was hospitalized on (b)(6)2021 for an allergic reaction on the face.The recipient was discharged on (b)(6)2021.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained during stimulation.The device passed the electrical and mechanical tests performed.It is believed that the failure of this device is most likely due to a malfunction within the digital chip.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|