• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SAFESTEP HUBER NEEDLE SET 22G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SAFESTEP HUBER NEEDLE SET 22G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Device Problems Material Deformation (2976); Activation Problem (4042)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event Description
It was reported that the safety feature had difficulty to work when the needle was removed.There was no reported patient injury.No other information was provided.
 
Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of difficulty activating the safety mechanism was confirmed and the cause was determined to be use related.The product returned for evaluation was a one 22ga x 0.75" safestep safety infusion set.The returned product sample was evaluated and the needle was observed to be bent just distal of the non-advanced safety mechanism.The following observations were noted during the sample evaluation: usage residue was seen on the sample which proved that the product had experienced at least some use.The needle tip was barbed suggesting contact between the needle and port base.The safety mechanism was ultimately successfully activated; however, resistance was encountered when passing the safety mechanism collar over the bent region of the needle.Force applied at an angle to the needle axis, or if the needle is not inserted perpendicular into the port septum, can lead to bending of the needle shaft.It is advised to verify that the needle length is correct based on port reservoir depth, tissue thickness and the thickness of any dressing beneath the bend of the needle; if too long, needle and/or port may be damaged at insertion.Additionally, avoid excessive manipulation once the needle is in the port.No potential damage related to the manufacturing process was noted on the complaint sample.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event Description
It was reported that the safety feature had difficulty to work when the needle was removed.There was no reported patient injury.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 22G X 0.75IN WITH Y-SITE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10547642
MDR Text Key207359507
Report Number3006260740-2020-03283
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00801741066177
UDI-Public(01)00801741066177
Combination Product (y/n)N
PMA/PMN Number
K153440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberLH-0029YN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2020
Event Location Hospital
Date Manufacturer Received10/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-