MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 1 HAND 6X7CM CTN 100; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 4007
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/19/2020
Event Type  malfunction  
Event Description
It was reported that the dressing has no adhesion and fall off easily.No patient harm reported.As a backup was available, the procedure was completed with the replacement.
 
Manufacturer Narrative
We have now concluded our investigation into this complaint.A review of the batch manufacturing records could not be performed as lot number was provided incorrect, however, there are no indications to suggest that the device did not meet specifications upon release into distribution.A complaint history review was carried out using the part number provided.There have been further complaints reported with this failure mode in the past three years.The device was intended to be used for treatment, however, issues were experienced during application.As no samples were returned a product evaluation could not be carried out.Low adhesion may be caused by adhesion levels of the raw material used to make the film of this dressing.There are various controls in place during the manufacturing process of both the raw material and the dressing to identify when the adhesion is not in acceptable range, as per the product specification.There are also regular tests carried out to monitor the finished product.As with all adhesive products, the skin site must be thoroughly cleaned and dried prior to application.If there is any moisture on the skin surrounding the catheter, the adhesive performance of this dressing may be compromised.If the dressing becomes wet whilst applied, it may have to be removed and a new dressing applied.We have not been able to confirm a relationship between the event and the device.The root cause remains undetermined.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 1 HAND 6X7CM CTN 100
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key10560635
MDR Text Key207741048
Report Number8043484-2020-02912
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4007
Device Lot Number20190401
Date Manufacturer Received10/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-