Additional information was added to d10, h3, h4 and h6.H4: the lot was manufactured from march 11, 2020 - march 12, 2020.H10: the device was received for evaluation.The bag was marked in the lower left edge and the top right corner was cut where the customer likely drained the contents.Unaided visual inspection was performed which observed a tear at the lower left edge of the bag where the marking was observed.Functional testing was performed which revealed a leak on the marked location.Additional magnified inspection which observed a tear/hole on the location of the leak.The reported condition was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|