From the information provided, there is no indication that there was any device malfunction, nonconformance, or misuse that contributed to the reported event.Potential adverse events in the instructions for use (ifu) with the tablo system includes, but are not limited to, other, more serious, complications arising from dialysis, such as hemorrhage, air embolism, acidosis, alkalosis or hemolysis, can cause serious patient injury or death.Outset medical, inc.System engineer has reviewed site system logs with a procedure date of (b)(6) 2020.No related system alarms were found to have occurred during treatment.The device is functioning post treatment.A review of production records for this unit did not note any manufacturing nonconformances that would contribute to a product event.
|
It was reported that six hours into dialysis treatment, the patient, who was covid positive and was not a chronic dialysis patient expired during treatment.It was reported that the patient was elderly, and had multiple co-morbidities.The treating team attributed the event to the patient's pre-existing condition.
|