• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS CYSTO VIDEOSCOPE; CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS CYSTO VIDEOSCOPE; CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-240
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/17/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.Omsc reviewed the manufacturing history (dhr) of the subject device, and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair of the subject device at the service department of olympus keymed, it was found the instrument channel port of the subject device had been loosened.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.The exact cause of the reported phenomenon could not be conclusively determined.However, based upon the information from olympus keymed there was the possibility that the reported phenomenon was attributed to the excessive twisted stress being applied to instrument channel port and the parts around it due to the user handling.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS CYSTO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10571967
MDR Text Key208108655
Report Number8010047-2020-06763
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCYF-240
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-