Model Number 71674041
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/24/2020
Event Type  malfunction  
Event Description
It was reported that anti rotation bar broke outside the patient during surgery.The procedure finished with a smith and nephew backup device.Patient injuries were not reported.No surgical delay.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3: h6: the device, used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device broke and has signs of wear and tear from use.The device was manufactured in 2012.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of complaint history for the listed part revealed no prior complaints at this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will be issued for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTI ROTATION BAR
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10573495
MDR Text Key208131884
Report Number1020279-2020-04820
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010605238
UDI-Public03596010605238
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674041
Device Catalogue Number71674041
Device Lot Number12CSP0005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2020
Date Manufacturer Received11/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-