• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY FUSION OMNI-TOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY FUSION OMNI-TOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Catalog Number FS-OMNI
Device Problem Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter occupation: unknown.Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the cutting wire securing component located near the distal end of the sphincterotome has disconnected from the catheter.The cutting wire is intact, and remains securely attached to the sphincterotome at the proximal end.However, due to the catheter and securing component disconnection, the distal end of the cutting wire is no longer connected to the sphincterotome catheter at the distal end.The securing component has a longer section measuring 3.0 mm and a shorter section measuring 2.0 mm, therefore no part of the device is missing.Liquid was also observed in the catheter.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user: "do not over flex or bow tip beyond 90 degrees, as this may damage or cause cutting wire to break." other factors that can contribute to separation of the cutting wire securing component, and the catheter include manipulating the handle with the catheter in a coiled position or with the pre-curved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove pre-curved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or pre-curved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all fusion omni-tomes are subjected to a visual inspection, and functional test to ensure device integrity.The functional test includes bowing the sphincterotome to ensure the distal end responds to handle manipulation.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook fusion omni-tome.The physician detected the cutting wire was broken during the procedure.This event was not reportable.On (b)(6) 2020, the customer noted the cutting wire securing component disconnected at the patient end of device and detached in the patient.The cutting wire securing competent was removed by forceps.The device was received on 08-sep-2020 with the anchor separated from the catheter, but was not detached from the device.On (b)(6) 2020, clarification was received from the customer noting that the information was remembered incorrectly and no piece had detached in the patient.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION OMNI-TOME
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key10573577
MDR Text Key208180074
Report Number1037905-2020-00402
Device Sequence Number1
Product Code KNS
UDI-Device Identifier10827002319032
UDI-Public(01)10827002319032(17)230203(10)W4315386
Combination Product (y/n)N
PMA/PMN Number
K052051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/03/2023
Device Catalogue NumberFS-OMNI
Device Lot NumberW4315386
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Date Manufacturer Received08/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ELECTROTOME SURGICAL GENERATOR, UNKNOWN MODEL; OLYMPUS 260V, ENDOSCOPE
Patient Age70 YR
Patient Weight75
-
-