Date of event was approximated to (b)(6) 2018 based on aware date and occurrences of previous referenced events.Device evaluated by manufacturer: returned product to boston scientific consisted of an imager diagnostic catheter in the sterile pouch.It was reported that the device tip was broken intentionally by the customer.The device was examined for damage, or any irregularities.The device showed multiple areas of bends/kinks.The kinks that were noticed on the device were due to the shipment of the device in the folded manner.It was noticed inside of the sterile pouch that the tip was separated from the shaft.Inspection of the remainder of the device, apart from the observed damage revealed no damage, or irregularities.The complaint stated that the physician broke the tip to prove the fragility of the tip, however, it is not possible to determine how much force the physician used to break the tip, and no way of determining how that would compare to the device design force specification.
|