It was reported that the biomed stated the device sent down from the floor because the hypothermia setting was not working, the device would not cool and it was not reading the pad temperature.
Mss transferred to the technical support.
Per follow up 1 on 10sep2020 via biomed, they left a voicemail message requesting the additional information.
Per follow up 2 on 11sep2020 via biomed, they run the calibration on the device and passed the calibration.
They also filled the device and it was placed back in service.
|