WRIGHTS LANE SYNTHES USA PRODUCTS LLC 90 DEGREE SCREWDRIVER T-HANDLE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number SD03.505.005
Device Problems Break (1069); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Reporter is company representative.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2020, an inventory check occurred and multiple devices including a mesh cutter, right angle bender, plate cutter, 90 degree screwdriver t-handle, handle for 90° screwdriver, shaft for 90° screwdriver, and matrixrib screwdriver blade self-retaining f/90° screwdriver were found with malfunctions.When tested, the devices would not work properly; would not turn, hold screws, and some were stripped.Some instruments were bent or dull.There was no patient or procedure involvement.(b)(4).
 
Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6- visual inspection results part number: sd03.505.005 lot number h057946 product was received missing one of the two pin components (sd03.505.005.05).The remining pin was also bent/deformed.The hole in the sd03.505.005.01 90 degree screwdriver body where the pin was missing was deformed (oval-shaped) a dent was observed on the ø7.3 face of the 90 degree screwdriver body (sd03.505.005.01) near the location of the missing pin component.The complaint conditions of "broken" and "deformed/bent" was confirmed, as it matches the reported condition, but upon further analysis it was determined not to be related to the manufacturing process for the following reason: the product passed all inspection criteria (100% visual and ctq features) and was found to be within specifications at the time it was released to the warehouse.The certificate of compliance from supplier isimac certifies that all product was manufactured in accordance with the product drawing, and no nonconformance's related to the complaint condition were present for the complainant lot.During the investigation no unidentified product design/manufacturing issues or discrepancies were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities." device history lot part # sd03.505.005 synthese lot # h057946 supplier lot # na release to warehouse date: 09 aug 2016 manufactured by synthes brandywine this non-conformance is not relevant to the complaint condition since the non-reworkable device was scrapped.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
90 DEGREE SCREWDRIVER T-HANDLE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10581669
MDR Text Key208384178
Report Number2939274-2020-04317
Device Sequence Number1
Product Code HWE
UDI-Device Identifier10887587071671
UDI-Public(01)10887587071671
Combination Product (y/n)N
PMA/PMN Number
K082649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSD03.505.005
Device Catalogue NumberSD03.505.005
Device Lot NumberH057946
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2020
Date Manufacturer Received10/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-