MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SINGLE-INNER SETSCREW; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179702000
Device Problem Device Slipped (1584)
Patient Problems Failure of Implant (1924); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Additional procode: kwp;kwq;mnh;mni;osh.Complainant part is not expected to be returned for manufacturer review/investigation.(b)(4).Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Investigation summary complaint summary: it was reported that on an unknown date, a female patient stated felt a ¿clunking¿ in the left side of her posterior lumbar spinal fusion surgery after her surgery on (b)(6) 2020.She received a follow up ct scan to show that the rod on her left l3 vertebrae has migrated but was still captured in the tulip.Currently, no other systems and no plan for revision surgery.Patient and procedure outcome were unknown.The implant(s) was not returned and instead the investigation will be done based on the supplied image(s).The image(s) was reviewed and the complaint condition could not be confirmed.As the implant(s) was not returned an as received condition, dimensional inspection, material or drawing reviews are not applicable.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, a female patient stated she felt a ¿clunking¿ in the left side of her posterior lumbar spinal fusion surgery after her surgery on (b)(6) 2020.She received a follow up ct scan that showed that the rod on her left l3 vertebrae has migrated but was still captured in the tulip.Currently, no other systems and no plan for revision surgery.Patient and procedure outcome were unknown.Concomitant device reported: unknown pedicle screw (part# unknown, lot# unknown, quantity unknown).This report is for one (1) single-inner setscrew.This is report 1 of 2 for (b)(4).
 
Event Description
Concomitant device reported: unknown screw (part# unknown, lot# unknown, quantity 3), unknown rod (part# unknown, lot# unknown, quantity 1).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE-INNER SETSCREW
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10586828
MDR Text Key208551632
Report Number1526439-2020-01794
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034064307
UDI-Public(01)10705034064307
Combination Product (y/n)N
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179702000
Device Catalogue Number179702000
Was Device Available for Evaluation? No
Date Manufacturer Received09/29/2020
Patient Sequence Number1
Treatment
UNKNOWN RODS; UNKNOWN RODS; UNKNOWN SCREWS; UNKNOWN SCREWS; UNKNOWN RODS; UNKNOWN SCREWS
Patient Outcome(s) Required Intervention;
-
-