• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PATELLA RESECTION GUIDE; PRST,KN,PAT/FEM,SEMCNSTR,UNCM,POR,COAT,POL/MET/POL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PATELLA RESECTION GUIDE; PRST,KN,PAT/FEM,SEMCNSTR,UNCM,POR,COAT,POL/MET/POL Back to Search Results
Model Number 71440391
Device Problems Component Missing (2306); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Patient Involvement (2645)
Event Date 08/31/2020
Event Type  malfunction  
Event Description
It was reported that during inspection, patella resection guide was noticed to have a screw missing.
 
Manufacturer Narrative
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the screw is missing from the patella resection guide.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATELLA RESECTION GUIDE
Type of Device
PRST,KN,PAT/FEM,SEMCNSTR,UNCM,POR,COAT,POL/MET/POL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10591565
MDR Text Key208711500
Report Number1020279-2020-04902
Device Sequence Number1
Product Code MBH
UDI-Device Identifier03596010453945
UDI-Public03596010453945
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440391
Device Catalogue Number71440391
Device Lot Number16FSP0022
Date Manufacturer Received09/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-