Model Number KD-V411M-0720
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The cutting wire of the subject device was broken.The broken section of the cutting wire was melted and burned.The coating on the cutting wire was partially torn.As a measurement result of the outer diameter of the cutting wire, there was no abnormality.As a measurement result of the length of the cutting wire and the cutting wire coating, there was no abnormality.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, omsc assumes that the damage of the coating occurred due to contacting with the metal part of the forceps elevator of the endoscope.The exposed cutting wire from the damaged coating contacted to the metal part of the forceps elevator while the device was activated, and it caused spark.Then, a part of the cutting wire became extremely hot, resulting in breakage.The above device handling has warned in the instruction manual as follows.Since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the wire is very short, the output is too high or activated while the wire touches metal parts of the endoscope, or the wire is tightened too strong.When the wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the wire will be pushed out toward the papilla or move sideways.If the wire breaks off, stop the output immediately and pull the slider completely to retract the broken wire into the tube.Then withdraw the instrument from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct and/or damage of the endoscope could result.When inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.Do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.
 
Event Description
It was reported that the cutting wire of the subject device broke during an inspection before use.There was no patient injury reported.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.On (b)(6) 2020, omsc found that the cutting wire was melted and burned.Therefore, we considered it was broken off during energization.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME V
Type of Device
SINGLE USE 2-LUMEN SPHINCTEROTOME
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10598021
MDR Text Key214151722
Report Number8010047-2020-06954
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K950166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKD-V411M-0720
Device Lot Number05K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2020
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-