SMITH & NEPHEW, INC. ASCOPE ACLAVE HD 4MM X 30 DEG; ARTHROSCOPE Back to Search Results
Model Number 72202087
Device Problem Material Separation (1562)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Event Description
It was reported that the tip lens of the scope was cracked.No case involved.Therefore, there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3,h6: the device, which was not used in a procedure, was returned for evaluation.There was a relationship between the reported event and the device.A visual inspection was performed and showed the scope to have severe distal tip and fiber damage, a damaged sidearm and a scratched distal lens.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.This failure is confirmed not originate from manufacturing, packaging, or labeling defects.This damage is caused by contact with another source.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.No manufacturing related defects were observed.No further investigation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCOPE ACLAVE HD 4MM X 30 DEG
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key10603025
MDR Text Key209175466
Report Number3003604053-2020-00113
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010616869
UDI-Public03596010616869
Combination Product (y/n)N
PMA/PMN Number
K962075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202087
Device Catalogue Number72202087
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Date Manufacturer Received10/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-