MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA PRO XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA PRO XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S40PRO
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus repair center for evaluation.The evaluation of the device the following: defects of the power supply unit and ignition coil were noted.The reported event was caused by these defects.If additional information becomes available, this report will be supplemented.
 
Event Description
During procedure, main lamp of the device did not ignite and spare lamp worked.The procedure was completed.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of the reported event could not be conclusively determined.However, omsc assumed that the defects of the power supply unit and ignition coil were caused by aging.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA PRO XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10604004
MDR Text Key209186328
Report Number8010047-2020-06992
Device Sequence Number1
Product Code GCT
UDI-Device Identifier04953170228940
UDI-Public04953170228940
Combination Product (y/n)N
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Type of Report Initial,Followup
Report Date 11/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-S40PRO
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-