Model Number CLV-S40PRO |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/26/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus repair center for evaluation.The evaluation of the device the following: defects of the power supply unit and ignition coil were noted.The reported event was caused by these defects.If additional information becomes available, this report will be supplemented.
|
|
Event Description
|
During procedure, main lamp of the device did not ignite and spare lamp worked.The procedure was completed.There was no report of patient injury associated with this event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of the reported event could not be conclusively determined.However, omsc assumed that the defects of the power supply unit and ignition coil were caused by aging.If additional information becomes available, this report will be supplemented.
|
|