Model Number SONATA |
Device Problem
Output Problem (3005)
|
Patient Problem
Failure of Implant (1924)
|
Event Type
Injury
|
Manufacturer Narrative
|
The device has not been explanted.If it should be explanted, it is to be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
|
Event Description
|
The user's hearing performance with the device started to gradually decrease a year ago.
|
|
Event Description
|
The user's hearing performance with the device started to gradually decrease a year ago.An incident of trauma or infection was not reported.Re-implantation is considered, however, a date for surgery has not been scheduled yet.
|
|
Manufacturer Narrative
|
Additional information: according to the currently available information, damage to the active electrode likely caused by minute device mobility is suspected.However to determine an exact root cause a device investigation of the explanted device is necessary.Re-implantation is considered but no date has been scheduled yet.
|
|
Event Description
|
The user's hearing performance with the device started to gradually decrease a year ago.An incident of trauma or infection was not reported.The user was re-implanted.
|
|
Manufacturer Narrative
|
Conclusion: damage to the active electrode which is consistent with minute device mobility was determined to have led to device failure over time due to fatigue wire breakages.The problems given in the recipient report appear to match the damage found.This is a final report.
|
|