OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE ELECTROSURGICAL KNIFE Back to Search Results
Model Number KD-611L
Device Problem Insufficient Information (3190)
Patient Problem Injury (2348)
Event Type  Injury  
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp (omsc). The malfunction of the subject device concerning this case has not been reported. The exact cause could not be determined. If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp. (omsc) received a literature titled "peroral endoscopic myotomy in treatment-naïve achalasia patients versus prior treatment failure cases". The literature reported the result of 502 achalasia patients with peroral endoscopic myotomy (poem) using olympus gif-h190 and kd-611l or kd-640l or kd-645l and fujifilm dh-28gr and erbe vio300d from (b)(6) 2013 to (b)(6) 2016. In the subject cases, enlargement of mucosal incision occurred in 4 patients. Therefore, omsc will submit a medical device report (mdr) depending on the event type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSINGLE USE ELECTROSURGICAL KNIFE
Type of DeviceSINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10605977
MDR Text Key215577993
Report Number8010047-2020-07021
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K092309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,literature
Reporter Occupation
Type of Report Initial
Report Date 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model NumberKD-611L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data