• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problem Output Problem (3005)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2019
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted as part of a retrospective review associated with the cfx longevity estimator software error advisory (report/fda recall number 2182208-10-02-2019-004-c).Concomitant medical products: 6721m-50, lead, implanted (b)(6) 2018; 6721m-50, lead, implanted (b)(6) 2018; 4968-60, lead, implanted (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited shorter than expected longevity estimate due to a programmer software estimator error.The programmer remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10610010
MDR Text Key209367036
Report Number2182208-2020-02181
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/13/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0239-2020
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient Weight99
-
-