Model Number NGP680300
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.Title, first name, last name: unknown.(b)(6).The system was evaluated by a field service engineer.The field service engineer (fse) found a burnt switching power supply.The fse replaced the power supply module to resolve the issue.A field service checklist was performed.The unit complied with all factory settings.A record review was performed.A product deficiency review was performed and there is no product deficiency identified.A document, service history, and trending was reviewed.There is not a recognizable adverse trend.The risks and mitigation associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.A labeling review was conducted; the operator manual for the system was reviewed and found to include adequate instructions for use, warnings and operational errors.The review of the device history record (dhr) for the whitestar signature system showed that there were no issues or non-conformities.The system and its components met all specifications prior to being released.Manufacturing has been ruled out as a potential cause for the reported issue.Based on the investigation results, no corrective action has been issued.Based on the investigation results there is no indication of a product quality deficiency.Johnson & johnson surgical vision will continue to monitor this type of complaints.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
A surgery center reported there was a burnt switching power supply.No patient injury reported.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITESTAR SIGNATURE SYSTEM
Type of Device
PHACOFRAGMENTATION UNIT
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
Manufacturer (Section G)
JOHNSON & JOHNSON SURGICAL VISION, INC.
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
1700 east st. andrew place
santa ana, CA 92705
7142478200
MDR Report Key10610836
MDR Text Key209325120
Report Number3006695864-2020-00447
Device Sequence Number1
Product Code HQC
UDI-Device Identifier05050474534476
UDI-Public(01)05050474534476
Combination Product (y/n)Y
Reporter Country CodeCN
PMA/PMN Number
K060366
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial
Report Date 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2020
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNGP680300
Device Catalogue NumberNGP680300
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-