Model Number MMT-1780KPK
Device Problem Power Problem (3010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/21/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Information received by medtronic indicated that insulin pump received open book image error.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device received with critical pump error due to moisture damage to force sensor.Device received with corroded electronic assemblies and corroded motor home switch.Misaligned force sensor cable and intermittent connection noted.Device unable to perform displacement test, rewind, prime/seating, basic occlusion, force sensor, occlusion, sleep current measurement, active current measurement, and self test due to critical pump error (open book image) alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1780KPK 670G PATHWAY BLACK MG
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key10611100
MDR Text Key209341290
Report Number2032227-2020-184831
Device Sequence Number1
Product Code OZP
UDI-Device Identifier000000763000283520
UDI-Public(01)000000763000283520
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1780KPK
Device Catalogue NumberMMT-1780KPK
Device Lot NumberHG45BRZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Is the Reporter a Health Professional? No
Date Manufacturer Received11/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/03/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age42 YR
-
-