It was reported that the arctic sun device received an alert 52 (extended period of cold water).
The cooling was started 6 hours ago.
The target temperature was 33c.
The patient temperature was 36.
1c at the start of the call by the end it was 36c.
The water temperature was 5c and the flow rate was 3.
1 l/m.
The 5 pads were in good coverage.
The esophageal probe in place.
The second temperature source correlating and the bair hugger in place, according to the protocol.
On a low setting no shivering was noted, according to the nurse.
The trend was in middle at the start of the call, 1 bar down.
The nurse checked the patient's skin and appears to be fine.
The nurse explained the machine was doing what it should be.
The nurse will speak to the doctor regarding the infection and possibly will start the (b)(4).
Ms&s instructed the nurse to call back when the alarm goes off again or for any other issues.
The nurse was aware of the possible skin issues when the water was too low for too long.
|