Model Number U125
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem Bradycardia (1751)
Event Date 08/03/2020
Event Type  malfunction  
Manufacturer Narrative
Additional information is being requested from the field. If additional information is received this report will be updated.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) and non-boston scientific right ventricular (rv) lead exhibited pace impedance measurements of greater than 2000 ohms, triggering a lead safety switch (lss). There was asystole of greater than two seconds due to myopotential oversensing. The field representative had programmed the device back to bipolar. The noise and high impedances could not be reproduced in the office. Technical services (ts) recommended programming options. The device remains in service. No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVALITUDE CRT-P
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key10620939
MDR Text Key209615351
Report Number2124215-2020-17566
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/16/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date10/03/2018
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number703896
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/31/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-