Model Number U125
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem Dizziness (2194)
Event Date 08/11/2020
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) and non boston scientific right ventricular (rv) lead exhibited impedance measurements of greater than 2000 ohms, noise, oversensing, and pacing inhibition of greater than two seconds.The patient experienced lightheadedness.The health care professional (hcp) reprogrammed the device and the device remains in service.No adverse patient effects were reported.
 
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device's spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) and non boston scientific right ventricular (rv) lead exhibited impedance measurements of greater than 2000 ohms, noise, oversensing, and pacing inhibition of greater than two seconds.The patient experienced lightheadedness.The health care professional (hcp) reprogrammed the device and the device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key10621070
MDR Text Key209618674
Report Number2124215-2020-17534
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public00802526559389
Combination Product (y/n)N
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/26/2020
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number710010
Was Device Available for Evaluation? No
Date Manufacturer Received09/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age73 YR
-
-