Model Number V173
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problems No Code Available (3191); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/18/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) reverted to safety mode unexpectedly.Technical services (ts) discussed potential causes and recommended device replacement due to limited therapy available in safety mode.Surgical intervention was undertaken.The device was explanted and replaced.No additional adverse patient effects were reported.The product has been received for analysis.This report will be updated upon completion of analysis.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) reverted to safety mode unexpectedly.Technical services (ts) discussed potential causes and recommended device replacement due to limited therapy available in safety mode.Surgical intervention was undertaken.The device was explanted and replaced.No additional adverse patient effects were reported.The product has been received for analysis.This report will be updated upon completion of analysis.
 
Manufacturer Narrative
F10: patient code 3191 captures the reportable event of surgery.Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.The system resets caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key10621752
MDR Text Key209640233
Report Number2124215-2020-17959
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public00802526536632
Combination Product (y/n)N
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/01/2015
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number103885
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2020
Date Manufacturer Received08/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-