• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938740
Device Problem Difficult to Open or Close (2921)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The event occurred on unspecified date of (b)(6) 2020.(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that ten (10) 2000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bags had issues with the clamps that will not close and when they do close, they then pop off.This was identified prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6.H4: the lot was manufactured from may 22, 2020 - may 23, 2020.H10: the actual samples were not available and therefore could not be evaluated; however, one (1) companion sample was received for evaluation.Unaided visual inspection did not identify any abnormalities that could have contributed to the reported condition.The clamp was evaluated which observed that it took excessive force to snap and shut the clamp and the clamp remained closed.The reported condition was verified.The cause of the condition is a manufacturing related issue.This issue is being further investigated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10623477
MDR Text Key209771645
Report Number1416980-2020-06167
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477244
UDI-Public(01)00085412477244
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Catalogue NumberH938740
Device Lot Number60241405
Date Manufacturer Received10/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-