This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be iled as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported by the sales rep via phone that during an unknown procedure the vapr 3 footswitch was shorting out.The complaint device was received and evaluated.Visual observations reveal that the cord was defective due to the cables were exposed.In addition, the device presented several marks of wear.Besides, both pedals presented signs of corrosion and the black push button is missing.The functional test was not performed due to the damages in the cord.The complaint can be confirmed.This device is 11 years old and its failure can be attributed to normal field wear.However, this cannot be conclusive determined.A manufacturing record evaluation was performed for the finished device lot number:0904038, and no non-conformances were identified.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|