Model Number 71364451
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/08/2020
Event Type  malfunction  
Event Description
It was reported that item broke in surgery.It was outside the patient.The procedure was completed without delay using an smith&nephew backup device.Surgery was not delayed.Patient was not harmed.
 
Manufacturer Narrative
The device, intended for use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 DEPTH GAUGE
Type of Device
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10624640
MDR Text Key209752682
Report Number1020279-2020-05124
Device Sequence Number1
Product Code KTT
UDI-Device Identifier03596010599438
UDI-Public03596010599438
Combination Product (y/n)N
PMA/PMN Number
K093047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364451
Device Catalogue Number71364451
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received03/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-