BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U35130126
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Injury (2348)
Event Date 06/16/2020
Event Type  Injury  
Manufacturer Narrative
A lot history review was conducted and it was determined that a device history record review was not required. The investigation is inconclusive for the reported detachment issue, as the device was not returned for evaluation. The definitive root cause for the reported detachment issue could not be determined based upon available information. The review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate. (expiry date: 09/2022).
 
Event Description
It was reported that during an angioplasty procedure, the device allegedly had a detachment issue. Reportedly, surgeon has been performed cut-down to access the balloon and remove. The current patient status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameULTRAVERSE 035 PTA CATHETER
Type of DevicePTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI N A
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key10625052
MDR Text Key209761481
Report Number2020394-2020-05920
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberU35130126
Device Catalogue NumberU35130126
Device Lot NumberCMDX0395
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/23/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/05/2020 Patient Sequence Number: 1
-
-