• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 630G INSULIN PUMP MMT-1715KL 630G; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hyperglycemia (1905); Blood Loss (2597)
Event Date 03/17/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Unable to perform the displacement test and the p-cap, reservoir will not lock properly due to missing retainer.Insulin pump received with missing reservoir tube o ring, and fading serial number label.Unable to performed occlusion test due to missing retainer.
 
Event Description
Information received by medtronic indicated that the plastic part of the introducer needle came away and needle stayed in the skin.Customer reported that customer experienced high blood glucose level and treated with bolus.When customer was removed infusion set and site was bleeding.Customer did not received medical treatment.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
8185464805
MDR Report Key10625098
MDR Text Key209769424
Report Number2032227-2020-185787
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00763000166519
UDI-Public(01)00763000166519(17)220529
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 10/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/29/2022
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberHG3FZQ7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2020
Date Manufacturer Received09/24/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age50 YR
Patient Weight82
-
-