Model Number N/A
Device Problems Fracture (1260); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/05/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Concomitant medical devices: 31-301301-arcos con sz a std 60mm trl-464320; 11-300818-arcos 18x150mm spl tpr dist-104510.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the screw driver tip broke inside the arcos cone body stem and they were unable to separate them apart.There was no patient harm or delay.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.One arcos 3.5mm hex drive and one arcos con sz a std 60mm trl were returned and evaluated.Upon visual inspection the driver had fractured with the fractured tip remaining inside of the taper body.There are wear lines on the shaft of the driver from us.A review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS 3.5MM HEX DRIVE
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10627293
MDR Text Key209799610
Report Number0001825034-2020-03705
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 01/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number31-301852
Device Lot NumberZB170704
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2020
Was the Report Sent to FDA? No
Date Manufacturer Received01/04/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNI
Patient Sequence Number1
Treatment
SEE H10
-
-