H4: device manufactured between may 30, 2020 to june 04, 2020.H10: only one (1) actual sample was received for evaluation, therefore, the other actual sample could not be evaluated.Additionally one (1) companion sample was received and evaluated.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed which revealed a leak at the spike port bonding area in back of the bag with the actual sample and the companion sample.The leak was observed between the spike port tube and the spike port bonding area in back of the bag with both samples.The reported condition was verified.The cause was not determined; however the most likely cause was due to inadequate or lack of cyclohexanone being applied to the cap tubing when it was inserted to the spike port during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|