Model Number CRPLUS
Device Problem No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/11/2020
Event Type  malfunction  
Manufacturer Narrative
Physio-control provided the customer with technical assistance and advised the customer to remove their device from service and obtain a replacement, as the device is not field serviceable and no long under warranty. The device was not returned to physio-control for evaluation. The cause of the reported issue could not be determined.
 
Event Description
The customer contacted physio-control to report that their device would not provide any voice prompts. Without voice prompts, the user would not able to use the device properly on a patient if needed. There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIFEPAK CR(R) PLUS DEFIBRILLATOR
Type of DeviceAUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key10632147
MDR Text Key210204010
Report Number0003015876-2020-01427
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K033275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 10/05/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberCRPLUS
Device Catalogue Number99403-000157
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-