SMITH & NEPHEW, INC. REDAPT DISTAL REAMER 190MM SZ 13 PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALIC Back to Search Results
Model Number 71355344
Device Problems Defective Device (2588); Material Twisted/Bent (2981)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/08/2020
Event Type  Injury  
Event Description
It was reported that the surgeon reamed femoral canal with the redapt 13x190 distal reamer. When the time came to remove, the reamer would not stay connected to the adaptor. There was a delay of roughly 45 minutes as surgeon tried to retrieve the reamer from the patient. Once removed from the patient and after inspection, the proximal end of the reamer was found to be bent. No backup s&n instrument was available and procedure was completed with the same device. No patient injury occurred. All available information has been disclosed. If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREDAPT DISTAL REAMER 190MM SZ 13
Type of DevicePROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key10632326
MDR Text Key209945524
Report Number1020279-2020-05164
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 10/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model Number71355344
Device Catalogue Number71355344
Device Lot Number16AUO0017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/15/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 10/05/2020 Patient Sequence Number: 1
-
-