• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-00
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/08/2020
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).Review of the dhr did not identify any deviations or nonconformities relevant to the issue.The affected device was sent to the manufacturer site for investigation.Results confirmed the reported issue and traced it back to a defective component on the motor control board.The part was replaced, and the issue solved.Subsequent functional verification testing was completed without further issues, and the unit was returned to service.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp displayed an error code associated to discrepancies between the set, and the actual rpm value during procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10634645
MDR Text Key210219447
Report Number9611109-2020-00557
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeSI
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-